STYROS® 2R

STYROS® 2R (Reversed Phase Polymeric)

 Key Features:

Feature 1:
High protein capacity (measured with Lysozyme)
10 mg/ml

 Feature 2:
Although we have used pressures of up to 10,000 psi without irreversibly altering the structure of the bed, we recommend the following pressure limits:
3,000 psi for the XH series,
3,000 psi for the NB series.
3,000 psi for the MB series
2,000 psi for the Micro Trap series
2,000 psi for Silica Capillary columns

Feature 3:
No measurable shrinking and/or swelling due to highly cross linked nature of the stationary phase.

Specifications:

Support Matrix

Cross linked Poly(styrene-divinylbenzene)

Surface Functionalities

Phenyl/Aryl

Shipping Solvent

Acetonitrile/Water 95/5 (v/v)

Packing Density

0.35 g/ml

Shrinkage/Swelling

< 0.5 % from 0 to 100 water to organic

Maximum Pressure Tolerance

207 bar (3,000 psi, 21 MPa)

 

These columns are offered in many sizes in PEEK and Stainless Steel